H2S H2O CO2 Online Process Analyzer – XLE 106-1-2-4


3-in-1 Gas Monitoring Made Simple

Choose Texas Analytical Controls for reliable, economical H2S, H2O, and CO2 monitoring in a single unit: the XLE106-1-2-4. With redundant H2S sensors, a capacitive H2O sensor for real-time moisture readouts, and an CO2 sensor (ranging up to 100%), this analyzer ensures consistent field monitoring over time. An intuitive touch screen controller allows you to easily navigate displays and adjust set-points.

Easy to Operate and Maintain

The XLE106-1-2-4 outshines the competition when it comes to ease of operation and maintenance. Sensors are easy to replace in just a few minutes without special tools, lasting one year for H2S/H2O and up to two years for CO2. Calibration is simple, taking less than six minutes, and can be set to occur automatically to save time in the field or be performed manually.

Full Technical Support

You will find the analyzers are easy to maintain. Technical support is available via phone or in person by one of our service technicians.

TAC XLE106-1-2-4 H2S, H2O and O2 Gas Analyzer

Downloads


XLE 106-1-2-4 Analyzer Spec Sheet PDF

Flow Diagram- H2S

Application Worksheet

XLE 106-1-2-4 Features


Front Loading Sensor

Minimize maintenance downtime. Our updated design allows operators to quickly and easily replace H2S sensors by removing just one thumbscrew.

H2S Analyzer quick change sensor panel with single thumb screw.

“Sensor Weak” Alert

Stay ahead of maintenance schedules. The analyzer automatically displays a “Sensor Weak” alert when H2S sensors reach 75% capacity, giving operators ample time to swap them out before failure.

Sensor Weak warning shown on H2S Analyzer screen

In-Line Bump Testing

Verify your readings without the hassle of removing the sample line. This quick check saves operators from having to perform a full, 10-minute calibration.

H2S Analyzer screen showing Bump Test in progress

On-Screen Historical Trend

Monitor performance at a glance. The display features a live, visual graph showing the historical trend of H2S readings over the previous hour.

H2S Analyzer historical trending chart displayed on-screen as a line graph

Automated “Flow Fail” Alerts

Ensure your sample is secure. The display immediately reads “Flow Fail” if regulators or flow meters malfunction, or if the sample line to the analyzer becomes obstructed.

Flow Fail displayed on screen

SD Card Data Logging & Custom IDs

Easily track and export your data. Store historical calibration and alarm data directly to an SD card. You can assign a custom Name or ID to each analyzer, which automatically carries over when downloading your data to a spreadsheet.

H2S Analyzer SD Card Holder location diagram

XLE 106-1-2-4 Specs


Range: Customer to specify a range and can be field programmable to alarm at any desired set point. (% level is also available)

  • H2S 0-10,25,50 and 100PPM. (% level is also available)
  • H2O 0-10, 0-20 LBS (other ranges available)
  • CO2 PPM and % Levels

Detection Method:

H2S- Electro-chemical (Estimated life: 1 Year)
H2O- H2O- Thermoset Polymer Sensor (Estimated life: 2 Years)
CO2- NDIR – Non-dispersive Infrared (Estimated life: 3 Years)


Calibration: An LED indicator on the front panel illuminates when the analyzer is in calibration mode. Display shows “Cal Complete” after a successful calibration. Calibration can be performed Automatically or Manually. Watch video below for details.


Communications Output:

4-20 mA (Self Powered)
Modbus
(See Customizations for additional options)


Sample Input Pressure: Inlet 10 +/-2 PSIG


Response Time: < 60 Seconds


Instrument Accuracy: +/-5% of the Full Scale


Instrument Repeatability: +/- 2%


Current Draw: < 1 A


Operating Temperature: -20°C to 50°C; -40°F to 122°F

XLE 106-1-2-4 Customizations


Range:

Standard

0-5 PPM
0-10 PPM
0-20 PPM
0-50 PPM
0-100 PPM

Medium

0-500 PPM
0-1000 PPM
0-3000 PPM

High

0-5000 PPM
0-10000 PPM

Percent

0-2%
0-5 %
0-10%
0-20 %

(Customer specifies desired range in PPM or %)


Communications Output:

4-20 Self Powered (Standard)
4-20 Loop Powered
1-5 VDC
Modbus via RTU (Standard)
Modbus via TCP/IP


Power Input:

12 VDC
24 VDC
110 VAC
Solar Panels


Alarm Relays:

Two Independent SPDT Relays for Low and High Alarm
Relays are set for FAIL SAFE
Alarm Setpoints, Latching, and Non Latching are PLC Programmable
Relay Ratings:
2.5A @ 125v AC
4.0A @ 24v DC


Alarm Delay: An Alarm Delay is PLC Programmable, the delay can be set for 0-90 minutes.


Shut-In Valve: Pneumatic solenoid valve or relays to control an external shut-in valve


Standard Enclosure: NEMA 4X powder coated (Dimensions: 12” x 10” x 6” Weight: 20 pounds)


Explosion Proof Enclosure: Explosion proof, suitable for Class 1, Division 1, Groups B, C, and D. (Dimensions: 18” x 16” x 12” Weight: 45 pounds)